• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Zimmer Biomet, Inc. AVENIR CMPL HA VAR COL SIZE 2; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Zimmer Biomet, Inc. AVENIR CMPL HA VAR COL SIZE 2; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer Back to Search Results
Catalog Number 574203020
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a primary total hip arthroplasty, a size 2 coxa vara femoral stem became firmly engaged distally during insertion and remained proud by approximately 1 to 2 cm, resulting in an intraoperative revision.During femoral preparation, the canal was rasped to size 2 and trialed with a coxa vara configuration, which was deemed to provide an appropriate fit.Upon implantation, the stem could not be advanced further despite multiple attempts.Extraction instruments were not available in theatre, necessitating an extended trochanteric osteotomy to facilitate removal.There was a significant intraoperative delay while additional instrumentation was transported from another site.Following removal of the stem, reconstruction was completed using an alternative stem and cone body.The patient underwent an extended trochanteric osteotomy and was reconstructed intraoperatively with an alternative stem and cone body.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).E1: (b)(6).G2: foreign - event occurred in united kingdom.H6: proposed component code : mechanical (g04) - stem.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVENIR CMPL HA VAR COL SIZE 2
Common Device Name
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer
Manufacturer (Section D)
Zimmer Biomet, Inc.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
Zimmer Biomet, Inc.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50725408
Report Number0001822565-2026-03091
Device Sequence Number18788373
Product Code LZO
UDI-Device Identifier00889024509276
UDI-Public(01)00889024509276(17)310407(10)3268945
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K182048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number574203020
Device Lot Number3268945
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-