• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Respironics, Inc. EverFlo; Generator, Oxygen, Portable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Respironics, Inc. EverFlo; Generator, Oxygen, Portable Back to Search Results
Model Number 1020006
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging no flow and burning smell.The device settings during the event were reported to be unknown.There was no allegation of serious or permanent harm or injury.No medical intervention was specified by the patient.Preliminary findings found that the clogged sieve caused the compressor to become overloaded and overheat, resulting in a burning smell.During the evaluation, the debug technician also noticed that the manifold was melted.Based on these findings, the sieves need to be replaced.Additionally, the inside of the rear cabinet is damaged, the whisper cap is cracked, and the capacitor is electrically defective.The inlet filter was replaced as part of preventive maintenance.The device was returned unrepaired to the manufacturer's product investigation lab (pil) for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EverFlo
Common Device Name
Generator, Oxygen, Portable
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50725444
Report Number2518422-2026-045418
Device Sequence Number10620863
Product Code CAW
UDI-Device Identifier00606959032552
UDI-Public(01)00606959032552
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K061261
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1020006
Device Catalogue Number1020006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-