• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Cordis US Corp. CATHETER EXTENSION; Set, Administration, Intravascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Cordis US Corp. CATHETER EXTENSION; Set, Administration, Intravascular Back to Search Results
Catalog Number 502102D
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2026
Type of Reportable Event Malfunction
Event or Problem Description
As reported, air bubbles formed immediately when a 100cm 1200psi angiographic catheter extension was connected to the device and operated, making it unusable.The same defect occurred in all five products.There was no reported patient injury.The device was not torqued against resistance while connecting to the tubing, and no difficulty was encountered while attempting to connect to a concomitant device.The procedure was continued using a 1200 mm braid tubing line.All defective products were discarded and were not retained.
 
Additional Manufacturer Narrative
As reported, air bubbles formed immediately when a 100cm 1200psi angiographic catheter extension was connected to the device and operated, making it unusable.The same defect occurred in all five products.There was no reported patient injury.The device was not torqued against resistance while connecting to the tubing, and no difficulty was encountered while attempting to connect to a concomitant device.The procedure was continued using a 1200 mm braid tubing line.All defective products were discarded and were not retained.The devices associated with complaints (b)(4) were not returned for product analysis and were manufactured from lot 18582463.A product history review (phr) of lot 18582463 was completed, and the review does not suggest that the failure experienced by the customer was related to the manufacturing process.The reported ¿catheter extension tubing - leakage¿ could not be substantiated because the devices were not returned for product analysis.The exact cause of the event cannot be determined based on the available complaint information.The complaint reported that the devices were not torqued against resistance during connection and that no difficulty was encountered while attempting to connect to a concomitant device.Review of the applicable angiographic catheter extensions eifu did not identify a specific warning or precaution addressing the reported leakage or formation of air bubbles during connection and operation.No evidence of use inconsistent with the product labeling or of a labeling-related contribution to the reported event was identified.Based on the available complaint information and manufacturing-history review, the reported leakage could not be substantiated, and its cause could not be established.Therefore, no further action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATHETER EXTENSION
Common Device Name
Set, Administration, Intravascular
Manufacturer (Section D)
Cordis US Corp.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
Cordis US Corp.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 32574
MX   32574
+011(786)3138372
MDR Report Key50725658
Report Number9616099-2026-01246
Device Sequence Number18788377
Product Code FPA
UDI-Device Identifier10705032064729
UDI-Public(01)10705032064729(17)290430(10)18582463
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K812231
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number502102D
Device Lot Number18582463
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-