H3: evaluation determined that it is not possible to perform a pre-temperature test due to the detached distal bearing.The likely cause of the failure could not be determined.It was also noted that the tip of the tube is worn out, the input and output drive shafts are worn and the laser markings are faded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|