• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Biosense Webster Inc VARIPULSE¿ Bi-Directional Catheter; Percutaneous Cardiac Abl Cath for Treatmt of AFib w Irreversible Electroporation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Biosense Webster Inc VARIPULSE¿ Bi-Directional Catheter; Percutaneous Cardiac Abl Cath for Treatmt of AFib w Irreversible Electroporation Back to Search Results
Catalog Number D141201
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Ischemic Heart Disease (2493); Pericarditis (4448)
Date of Event 05/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent pulmonary vein isolation and cavotricuspid isthmus (cti) ablation and the patient experienced pericarditis.Three days after the case, patient experienced acute breathlessness, and sore while lying on back.The patient went to "nhs".No pericardial effusion on echo, but symptoms consistent with pericarditis.It was also reported that there was an indication that st change on electrocardiogram may be a vagal response.On review of the case details from the carto system, 37 radiofrequency (rf) lesions were delivered at the cti with the thermocool smarttouch sf catheter.Pulmonary vein isolation was performed with the varipulse¿ bi-directional catheter plus workflow - 4 ablations per vein/16 total, with 4ml/min passive irrigation and 30ml/min during pulsed field ablation (pfa) application.The physician indicated initially that he thought the potential pericarditis could have been a result of pfa, although once he saw the number of rf lesions, he determined the cti ablations would be a more likely cause for any pericarditis/symptoms experienced by the patient in the days following the procedure.
 
Additional Manufacturer Narrative
D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.E1.Initial reporter phone: (b)(6).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VARIPULSE¿ Bi-Directional Catheter
Common Device Name
Percutaneous Cardiac Abl Cath for Treatmt of AFib w Irreversible Electroporation
Manufacturer (Section D)
Biosense Webster Inc
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
Biosense Webster Inc (Juarez)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
katie seppa
31 technology dr
irvine, CA 92618
+1(513)337-8052
MDR Report Key50726562
Report Number3013300026-2026-02983
Device Sequence Number18801281
Product Code QZI
UDI-Device Identifier10846835025460
UDI-Public10846835025460
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P240006
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberD141201
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age52 YR
Patient SexMale
-
-