It was reported that during use the cardiosave intra-aortic balloon pump (iabp) had a display distortion.Patient involved and no harm is reported.The issue recurred after restarting the device and as a precaution, the device was replaced and treatment continued.The issue discovered by the customer.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported display malfunctions (ecg, arterial pressure noise, and touchscreen).The relevant circuit boards have been cleaned and reconnected.During the repair, it was confirmed that a foreign object resembling a hair was found inside the upper display.As this was confirmed, and considering the possibility that it may have contributed to the current issue, the upper display was also replaced.
|