| Model Number |
LF1937 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Difficult to Open or Close (2921); Intermittent Energy Output (4025)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a vats esophagectomy procedure, at the time of dissection of the tissue surrounding the stomach, about 6 hours after starting use, it seemed that tissue had entered inside the jaw, it was difficult/unable to pull the cutting trigger, and the jaw could no longer open. another ligasure device was reported to have been used successfully with the same generator.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: vlft10gen, vlft10gen ft series energy platformx1, (sr unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|