| Model Number |
0998-00-3013-45 |
| Medical Device Problem Code |
Signal Artifact/Noise (1036)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during use, the cs100 intra-aortic balloon pump (iabp) had poor ecg signal when using the device.There was patient involvement.No harm was reported.Customer switched to another device the field service engineer reported that after replacing the ecg lead wires, the equipment functioned normally.Product passed all functional and safety tests per manufacturer specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|
Search Alerts/Recalls
|