Medtronic Xomed Inc. XPS® HANDPIECE - VISAO®; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I
|
Back to Search Results |
|
| Model Number |
3334800 |
| Medical Device Problem Code |
Vibration (1674)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during pre-op, the doctor installed the bur and performed a test.During operation, the motor produced an abnormal noise, the bur was wobbling.The doctor tried reinstalling the bur, but the issue persisted.Therefore, the doctor replaced the handpiece with a new one to complete the surgery.The same bur worked well with another handpiece.When other bur was tried with this handpiece, bur wobbled.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis found motor corroded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|