C.R. Bard, Inc. (BASD) -3006260740 Navarre Universal Drain; Chronic Catheters Back to Search Results
Catalog Number NNU6LPT
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2026, a patient underwent an ultrasound guided percutaneous transhepatic cholangiography drainage (ptcd) catheter placement procedure using a navarre universal drainage catheter.During the procedure, it was reported that leakage was observed from a small hole in the sure-lok component during drainage.The procedure was successfully completed using a replacement device.There was no reported patient injury.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent an ultrasound guided percutaneous transhepatic cholangiography drainage (ptcd) catheter placement procedure using a navarre universal drainage catheter.During the procedure, it was reported that leakage was observed from a small hole in the sure-lok component during drainage.The procedure was successfully completed using a replacement device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, two electronic photos were provided for review.The first and second photos show complete views of individual loaded drainage catheters.The third photo shows a complete view of two loaded drainage catheters.The distal end of both catheters noted to be in pigtail condition.The reported fluid leak cannot be verified from the provided photos.The investigation is inconclusive for reported fluid leak issue for both the devices as the provided photo was not sufficient to confirm the issue.The definitive root cause could not be determined based upon available information labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (method, result, conclusion).E1: (b)(6).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Navarre Universal Drain
Common Device Name
Chronic Catheters
Manufacturer (Section D)
C.R. Bard, Inc. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. Bard, Inc. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50730781
Report Number3006260740-2026-06632
Device Sequence Number10575167
Product Code GBX
UDI-Device Identifier00801741000348
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K951475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNNU6LPT
Device Lot NumberGFKN0713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Supplement Date Received by Manufacturer09/17/2026
Initial Report FDA Received Date09/11/2026
Supplement Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-