| Catalog Number |
NNU6LPT |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent an ultrasound guided percutaneous transhepatic cholangiography drainage (ptcd) catheter placement procedure using a navarre universal drainage catheter.During the procedure, it was reported that leakage was observed from a small hole in the sure-lok component during drainage.The procedure was successfully completed using a replacement device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent an ultrasound guided percutaneous transhepatic cholangiography drainage (ptcd) catheter placement procedure using a navarre universal drainage catheter.During the procedure, it was reported that leakage was observed from a small hole in the sure-lok component during drainage.The procedure was successfully completed using a replacement device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, two electronic photos were provided for review.The first and second photos show complete views of individual loaded drainage catheters.The third photo shows a complete view of two loaded drainage catheters.The distal end of both catheters noted to be in pigtail condition.The reported fluid leak cannot be verified from the provided photos.The investigation is inconclusive for reported fluid leak issue for both the devices as the provided photo was not sufficient to confirm the issue.The definitive root cause could not be determined based upon available information
labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (method, result, conclusion).E1: (b)(6).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|