The customer reported the last box purchased had about 5 individual gloves affected total, with fingertips all stuck together while the cuff was wadded and unfinished, another pair had 2-3 fingers that were not closed, and another looked unfinished; all unusable.Medline quality did not receive a sample or photograph of the reported item for investigation; a review of inspection record found no similar issues noted, a post-market retrospective review of the past 24 months preceding the reported issue found 0 additional complaints of gloves sticking together, and retained samples were reviewed and did not find any gloves that were stuck together or looked "unfinished." the issue is unconfirmed, as no sample nor photographs were made available to review, and the cause was documented as sample required to determine cause.Based on the findings of the quality investigation, no change has been made to the regulatory reportability decision for the complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|