| Model Number |
Mi1200 SYNCHRONY |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Clinic stated reason for explantation is device failure.The user was re-implanted.
|
| |
|
Event or Problem Description
|
|
Clinic stated reason for explantation is device failure.The user was re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode, as might be caused by an external mechanical impact, was determined to be the root cause of device failure.Other mechanical damages found are attributable to the removal surgery.This is a final report.
|
| |
|