A cranial surgery for a planned revision of a ventricular peritoneal shunt placement in the brain for drainage of cerebrospinal fluid (csf), due to shunt clogging with increased intracranial pressure, has been performed with the aid of the brainlab cranial navigation 4.1.After placement of the revision shunt, the surgeon received the expected csf and concluded the surgery.From a post-operative imaging scan, the surgeon determined that the placed revision shunt deviated from its intended position slightly medial and short of the ventricle, i.E.Not located within the ventricle as desired.The surgeon decided that a revision surgery is necessary to correct the shunt placement, the revision surgery was planned and took place on (b)(6) 2026.According to the surgeon (treating clinician):
- the surgery was a planned surgery, i.E.Not an emergency surgery.- there was no (direct or increased) risk of harm to a critical structure due to the deviating shunt placement.- there was no resulting harm or negative effect to the patient, except for that this planned revision shunt placement was not successful as intended.- there were no changes to the planned procedure at the surgery, and the surgery / anesthesia was not prolonged.- a revision surgery was planned to correct the shunt placement and took place on (b)(6) 2026.There was also no negative effect to the patient due to the repeat anesthesia/surgery of ca.2hrs.- there were further no remedial actions necessary, done or planned for this patient.Hospitalization was prolonged by one day for the patient returning for the period of the revision surgery.
|
|
B2, h1:
a risk to the patient's health could not be excluded for these specific circumstances, since a shunt catheter placement was performed in a different location in the brain than anticipated and desired with the brainlab navigation involved, despite according to the surgeon (treating clinician):
- the surgery was a planned surgery, i.E.Not an emergency surgery.- there was no (direct or increased) risk of harm to a critical structure due to the deviating shunt placement.- there was no resulting harm or negative effect to the patient, except for that this planned revision shunt placement was not successful as intended.- there were no changes to the planned procedure at the surgery, and the surgery / anesthesia was not prolonged.- a revision surgery was planned to correct the shunt placement and took place on (b)(6) 2026.There was also no negative effect to the patient due to the repeat anesthesia/surgery of ca.2hrs.- there were further no remedial actions necessary, done or planned for this patient.Hospitalization was prolonged by one day for the patient returning for the period of the revision surgery.H6:
according to the results of the brainlab investigation and the information provided by the hospital, it can be concluded that the root cause of the shunt placed in the brain with the aid of navigation deviating medial and short of the ventricle, missing its intended target, is:
- a not adequate distribution of surface registration points acquired by the user for the patient anatomy registration to the navigation / to the pre-operative ct scan, not following brainlab requirements as also guided by the brainlab software, in combination with a ct scan used for navigation registration that does not fulfill brainlab requirements.The navigation and imaging data from this surgery show that the registration points were not sufficiently distributed on the required distinctive (bony) surfaces, i.E.Not sufficiently on both sides of the patient's face and both sides of the nose - but in the front mainly only on rounded areas as the dome of the head.The image data set used cuts off under the eyes, not containing necessary areas to acquire points, additionally some points were collected in areas not available in the scan, which should be avoided and is impacting the registration calculation.This caused the navigation software to not find an as accurate match between the pre-operative image dataset and the actual patient anatomy in the region of interest, as desired for the catheter placement.Apparently, the resulting deviation between the actual patient anatomy location during the surgery and the registered pre-operative patient image scan, displayed by the navigation for instrument position visualization, was not recognized by the user with the required thorough navigation accuracy verification of the registration, and throughout the procedure, for e.G.After draping of the patient and before performing significant invasive actions.There is no indication of a systematic error or malfunction of the brainlab device (navigation).Corresponding brainlab measures to minimize this anticipated risk as low as reasonably practicable are already in place.H7:
brainlab intends to re-iterate the relevant topics regarding the use of the device to this customer.
|