The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review was unable to be performed as the lot number was not provided.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of urinary incontinence, hole/perforated, and mechanical issue was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of unable to exclude device problem was assigned to this investigation.If further information is received the investigation will be re-opened.This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation (ncep-00181844) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 126510003.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #:(b)(4).Model number/catalog number: 72400098.Serial number: n/a.Batch/lot number: 124876003.Model/catalog description: control pump fgs.Unique identifier (udi) #: (b)(4).
|