Investigation performed by r&d project manager in the absence of images and based solely on the description of the event, it is not possible to determine with certainty the causes that led to the pin failure.As these pins are used in nextar procedures, the instrumentation is equipped with specific guides to assist the surgeon during pin insertion.In particular, these guides are designed to ensure the correct trajectory of the pins, thereby reducing the risk of deviation in the event of insertion that is tangential to the cortical bone.Such deviation may subject the pin to bending forces, potentially resulting in breakage at the transition between the smooth and threaded portions of the pin.Root cause:as no lot number information or affected device could be provided, a detailed investigation could not be completed.While no definitive root cause can be established, the issue reported may have resulted from application-specific factors, such as excessive bending forces, possibly in combination with the unique patient-specific conditions associated with this case; however, this cannot be confirmed based on the available information.
|