The customer reported that the actuator is not working properly, the boom is not moving, and there is a clicking sound; the customer later reported that the replacement battery did not resolve the issue.Photo provided only shows the reference label of the lift, and product support did provide troubleshooting with the customer; after further troubleshooting with product support, it was determined that the actuator cable was not properly plugged into the control box.Investigation results: not confirmed.This complaint is due to customer knowledge, and after customer re-inserted the actuator cable into the control box, the lift is now functional; issue has been resolved through troubleshooting.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|