| Model Number |
24967 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Communication or Transmission Problem (2896); Problem with Software Installation (3013)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the blue portion of the patient connector used to hang around the patient's neck was damaged and required replacement.It was noted that the replacement connector displayed a security certificate expired message.It was further noted that the replacement connector could not be paired with the mobile programmer application due to bluetooth connectivity issues.Troubleshooting steps were taken to include removing the bluetooth pairing, rebooting the host tablet, and pairing the connector as a new device.It was further noted that a connector update needed message was displayed; however, the update failed despite multiple attempts. performance data collected from the mobile programmer was received and analyzed.Analysis determined that bluetooth connection was lost.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: performance data collected from the mobile programmer was received and analyzed.Analysis of the service logs found the customer comment that the replacement connector displayed a security certificate expired message and that the replacement connector could not be paired with the mobile programmer application due to bluetooth connectivity issues was confirmed.Analysis determined that bluetooth connection was lost.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|