| Model Number |
MX40 1.4 GHz Smart Hopping |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to the philips authorized repair facility for evaluation.Upon receipt of the device, it was visually inspected.Results of visual inspection identified the device had been tampered with and the rear housing case and front bezel display are not genuine philips parts.The product malfunction was unable to be confirmed because there was evidence the device had been opened and modified/repaired outside of a philips factory/repair bench, so the device was returned to the customer unrepaired.No further action is necessary at this time.
|
| |
|
Event or Problem Description
|
|
Philips received a complaint on mx40 1.4 ghz smart hopping indicating that there was a speaker malfunction error.It is unknown if the device produced sound.It is unknown if the device was in clinical use at the time of the event.No adverse event was reported.
|
| |
|