| Model Number |
407452 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Capturing Problem (2891); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that post implant the right ventricular (rv) lead dislodged and exhibited noise on stored episodes and high and rising thresholds.Oversensing and short v-v intervals were also noted.The lead was reprogrammed off with the device set to aai.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|