| Model Number |
D151 |
| Medical Device Problem Codes |
High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a defibrillation threshold (dft) test this implantable cardioverter defibrillator (icd) system recorded a code 1005, indicative of an open circuit condition.A commanded shock was delivered and high out of range shock impedance measurements greater than 125 ohms were detected.As a result, the right ventricular (rv) lead was capped and replaced, and this icd was explanted and replaced due to rv lead compatibility.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|