It was reported during right ventricular (rv) lead placement, while attempting to direct the delivery catheter to the target location, the proximal side at the catheter end bent and kinked, and proper manipulation and implantation became difficult.The catheter was attempted/not used and replaced.Furthermore, during slitting after ventricular lead placement, when the cutter reached the middle portion of the catheter, it did not advance smoothly, and this led to the lead being pulled and dislodgement occurring; therefore, the lead was placed again.The rv lead was successfully implanted. no patient complications have been reported as a result of this event.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|