The customer stated that the adhesive was really strong and that he felt his skin was going to peel when he was trying to remove the bandage; the bandage was applied to the right pinky finger for 3hrs, and the customer reported no allergic reaction or skin irritation.No physical or photo samples of the complaint issue were provided by the customer for review, limiting the ability to investigate the issue thoroughly; complaint data trending found no other complaints with this product regarding the adhesive being too strong, and the vendor's returned investigation found nothing within relevant documentation to indicate any potential contributors towards the complaint issue.Based on the information provided and the results of the investigation, medline was not able to confirm the reported issue or that the product failed to meet specifications.No corrective or preventative actions were implemented, and complaints are monitored via the trend analysis procedure.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|