| Model Number |
M635WU60310 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported a patient death occurred.A concomitant farapulse and left atrial appendage (laa) closure procedure was performed.Both cases went as planned and a 31mm watchman flx pro closure device was implanted.The patient stayed overnight and x-ray and echocardiogram showed the closure device in place and no pericardial effusion.The patient discharged home the following day.Six days later the patient was found deceased at their home.There is no further information at this time.
|
| |
|
Event or Problem Description
|
|
It was reported a patient death occurred.A concomitant farapulse and left atrial appendage (laa) closure procedure was performed.Both cases went as planned and a 31mm watchman flx pro closure device was implanted.The patient stayed overnight and x-ray and echocardiogram showed the closure device in place and no pericardial effusion.The patient discharged home the following day.Six days later the patient was found deceased at their home.There is no further information at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific (bsc) concludes:
device technical analysis:
the watchman flx? pro remains implanted; therefore, returned device analysis could not be completed.Device history record review:
a review was completed of the device history records (dhr) for the watchman flx? pro, part # m635wu60310 batch # 0037370621.The dhr review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review:
there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? pro instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include death.Risk review:
a risk review was completed and confirmed that the event of death was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|