| Catalog Number |
10445128 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that an erroneously depressed vancomycin (vanc) result on a patient sample was obtained on a dimension vista 500 system.The initial result was reported to the physician and questioned.The same sample was repeated on both original and alternate dimension vista 500 systems.The repeat results obtained were higher than the initial result, matched the clinical history and were considered correct.The repeat result from an alternate system was reported as the correct result to the physician.There are no known reports of patient intervention or adverse health consequences due to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
A united states (us) customer contacted the siemens customer care center (ccc) regarding an erroneously depressed vancomycin (vanc) result on a patient sample obtained on a dimension vista 500 system.Siemens reviewed the instrument data, and the information provided by the customer.Reagent and instrument issues were ruled out based on calibration and quality control (qc) recovery within acceptable ranges and no issues were reported with other patient samples.Based on the available information, the cause of the erroneous result was unable to be determined.The customer is operational.The device is performing according to specifications.No further evaluation is required.
|
| |
|