• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; 1LYR 16FR10ML 100%SILI UM TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; 1LYR 16FR10ML 100%SILI UM TRAY Back to Search Results
Catalog Number URO170816
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 1 ea syringe tip broken from lot: 23abr985.
 
Additional Manufacturer Narrative
The customer reported quality issues on uro170816, including 1 ea syringe tip broken.Medline quality did not receive a sample nor photographs of the reported product, and a retrospective review of the past 6 months preceding the reported issue revealed one additional complaint of broken syringe tips with respect to uro170816.The issue is unconfirmed, as no samples nor photographs were made available to review; based on the description of the issue, the root cause is identified as physical weakness at the site of the syringe tip and barrel.Medline quality will continue to monitor for the issue through weekly complaint meetings and monthly trending analysis, and if a product is received or more information is made available, the complaint will be reopened and an investigation will be performed.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
1LYR 16FR10ML 100%SILI UM TRAY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
joyce qiu
three lakes drive
northfield, IL 60093--275
MDR Report Key50733089
Report Number1417592-2026-36895
Device Sequence Number10591914
Product Code OHR
UDI-Device Identifier40888277862624
UDI-Public40888277862624
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberURO170816
Device Lot Number23ABR985
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2023
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1