| Model Number |
72400160 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this artificial urinary sphincter (aus) presented to the hospital with a non-functional device.Two weeks later, a surgical procedure was performed and upon inspection, there was a crack found in the cuff connecting tube.As a result, the physician removed and replaced the entire device.The patient was stable following the procedure, and there was no additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72400098.Serial number: n/a.Batch/lot number: 1100189697.Model/catalog description: control pump fgs.Unique identifier (udi) #:(b)(4).Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100220776.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).
|
| |
|