• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Xenios AG Novalung ultimate kit (US); Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Xenios AG Novalung ultimate kit (US); Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure Back to Search Results
Catalog Number F32000015
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
A user facility reported an underperforming oxygenator during extracorporeal membrane oxygenation (ecmo) support that required exchange in less than 8 hours, with a second oxygenator having the same reported issue in less than 12 hours.Upon internal review, the user facility stated that 10 of 16 oxygenators required change out in under one week due to failure.Three of these kits have been accounted for, however, this is to capture the remaining 7 kits included in the customer complaint.Data is still being collected from the customer for review.This case reflects an event where a patient on veno-venous ecmo support experienced suboptimal oxygenation with a post-oxygenator arterial oxygen of 118 mmhg on (b)(6) 2026 requiring product exchange.There was no indication of patient serious injury.The complaint sample was not available for product investigation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Novalung ultimate kit (US)
Common Device Name
Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure
Manufacturer (Section D)
Xenios AG
director of quality
im zukunftspark 1
heilbronn baden-wurttemberg 74076
GM  74076
Manufacturer (Section G)
Xenios AG
director of quality
im zukunftspark 1
heilbronn baden-wurttemberg 74076
GM   74076
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50733169
Report Number3012172416-2026-00076
Device Sequence Number10633061
Product Code QJZ
UDI-Device Identifier04057224001494
UDI-Public04057224001494
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K232926
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberF32000015
Device Lot NumberHSXA1906
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Novalung Console (Linux Board).
Patient SexUnknown
-
-