• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LUX-Dx II+?; Insertable Cardia Monitor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LUX-Dx II+?; Insertable Cardia Monitor Back to Search Results
Model Number M312
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an insertable cardiac monitor (icm) indicated an eos (end of service) condition during preparation for implant.The insertable cardiac monitor had been enrolled but was not implanted, and an alternate insertable cardiac monitor was used instead.Technical services requested return of the affected insertable cardiac monitor for evaluation; however, the hospital retained and subsequently discarded the insertable cardiac monitor, making product analysis unavailable.The representative later confirmed that the insertable cardiac monitor successfully transmitted a remote report prior to disposal.No patient impact occurred because the insertable cardiac monitor was not implanted.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that an insertable cardiac monitor (icm) indicated an eos (end of service) condition during preparation for implant.The insertable cardiac monitor had been enrolled but was not implanted, and an alternate insertable cardiac monitor was used instead.Technical services requested return of the affected insertable cardiac monitor for evaluation; however, the hospital retained and subsequently discarded the insertable cardiac monitor, making product analysis unavailable.The representative later confirmed that the insertable cardiac monitor successfully transmitted a remote report prior to disposal.No patient impact occurred because the insertable cardiac monitor was not implanted.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Device technical analysis: this product has not been returned to boston scientific and as a result, laboratory analysis could not be conducted.Device history record review: a review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUX-Dx II+?
Common Device Name
Insertable Cardia Monitor
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50733404
Report Number2124215-2026-48700
Device Sequence Number21498338
Product Code MXD
UDI-Device Identifier00802526620706
UDI-Public(01)00802526620706(17)271021
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K231328
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM312
Device Catalogue NumberM312
Device Lot Number189671
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2026
Initial Report FDA Received Date09/11/2026
Date Device Manufactured05/05/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age86 YR
Patient SexUnknown
-
-