Catalog Number ER320
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a laparoscopic cholecystectomy when loading the first clip, the trigger was locked, preventing the use of any clips on the device.Was there any harm to the reported patient or user? no.Was there a significant delay? no.
 
Additional Manufacturer Narrative
(b)(4).Date sent 9/11/2026.D4 batch #: z7050n.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the er320 device was returned with the jaws closed and with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality; however, the trigger could not be actuated due to a jammed condition.The device was subsequently disassembled to evaluate the condition of the internal components.Examination revealed that the trigger was bent, preventing the clips from advancing into the jaws.Additionally, sixteen (16) clips were found remaining within the clip track at the time of disassembly.Additionally, photographs were provided showing the condition of the sample as received.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch z7050n number, and no non-conformances were identified.Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGACLIP
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50733460
Report Number3005075853-2026-07471
Device Sequence Number16111927
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public(01)10705036012580(17)301130(10)842D76
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K812291
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberER320
Device Lot Number842D76
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/13/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-