Ethicon Endo-Surgery, LLC. LIGACLIP; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown surgery, after fired, could not open the jaw.Opened the jaw manually with big force.There were no adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent 9/11/2026.D4 batch #: a70028.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the er320 device was returned with a clip present in the jaws and no apparent damage to the external components.In addition, the opened packaging was returned along with the instrument.In an attempt to replicate the reported incident, the device was subjected to functional testing.During testing, the trigger could not be actuated due to a jammed condition.The device was subsequently disassembled to assess the condition of the internal components.Examination revealed that the trigger was bent, preventing clip advancement into the jaws.Additionally, fourteen (14) clips were found remaining within the clip track at the time of analysis.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch a70028 number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGACLIP
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50733641
Report Number3005075853-2026-07474
Device Sequence Number10662443
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public(01)10705036012580(17)310131(10)895D00
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K812291
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberER320
Device Lot Number895D00
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-