(b)(4).Date sent 9/11/2026.D4 batch #: a70028.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the er320 device was returned with no apparent damage.In addition, the opened packaging was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.During testing, the trigger could not be actuated as it was jammed.The device was subsequently disassembled to evaluate the condition of the internal components.Examination revealed that the trigger was broken, preventing the clips from advancing into the jaws.Additionally, seventeen (17) clips were found remaining within the clip track.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch a70028 number, and no non-conformances were identified.
|