| Model Number |
WA50040A |
| Medical Device Problem Codes |
Break (1069); Crack (1135); Poor Quality Image (1408); Dent in Material (2526); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a broken lens during reprocessing involving an olympus rigid video laparoscope.Additionally, during the initial evaluation of the device, a cracked objective lens, cracked light guide connector and cover glass, dents and scratches on the outer tube, and a cracked image were observed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided once available.The device was returned to olympus for inspection, and the reported failure was confirmed.Based on the results of the investigation a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|