Block h6: imdrf device code a050501 captures the reportable event of bands failure to deploy.Block h11: investigation results: the speedband superview super 7 device was disposed and not returned for analysis; therefore, a technical analysis could not be performed.A review of the device history record (dhr) was performed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A similar complaint review was completed.There were no emerging trends identified that warrant further action.A risk review was completed and confirmed that the event of "bands failure to deploy" was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.The labeling review was performed.The instructions for use (ifu) contains detailed device information and instructions for the device use.There is no evidence the device was improperly used per the ifu, and there is no evidence that there is any issue with translation, wording, or graphics of the ifu/labeling information.Based on the event description and information provided by the customer there is not enough evidence to determine whether the issue was induced due to the physician's technique during the procedure or was related to a device malfunction.Additionally, based on the most relevant information for the complaint including product record review results, the device met all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process, the definitive cause of the reported issue cannot be established due to lack of evidence.Without proper evaluation of the device, it remains unknown the causes that contributed to the event.Based on the thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|