Model Number D233
Medical Device Problem Code Failure to Capture (1081)
Health Effect - Clinical Code Bradycardia (1751)
Date of Event 08/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) called in due to interrogation issues with the home monitor.Technical services (ts) reviewed and found there was a recent transmission with no noted episodes.The tele monitor displayed there was a failure to capture and the rate dropped below the lower rate limit into the 40s.The rate went back to the 60s.This icd remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT? EL ICD DR
Common Device Name
Implantable Cardioverter Defibrillator (Non-Crt)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50734245
Report Number2124215-2026-48717
Device Sequence Number21487772
Product Code LWS
UDI-Device Identifier00802526587801
UDI-Public(01)00802526587801(17)271028
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberD233
Device Catalogue NumberD233
Device Lot Number719554
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2026
Initial Report FDA Received Date09/11/2026
Date Device Manufactured11/13/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
-
-