MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? Pro; system, appendage closure, left atrial

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? Pro; system, appendage closure, left atrial Back to Search Results
Model Number M635WU60240
Medical Device Problem Code Failure to Seal (4070)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that device leak occurred.\ \ a left atrial appendage (laa) closure procedure was performed, and a 24mm watchman flx pro closure device was implanted on (b)(6) 2026.During the procedure, there were multiple recaptures and repositions, the patient had a difficult anatomy and was a challenging procedure.The patient was discharged.During a routine follow up computed tomography (ct) scan, a 7mm leak was noted.The ct was sent to the therapy operations team for evaluation.Per the therapy operations report it stated that the distal portion of the appendage showed no signs of contrast nor contrast going around the closure device but that a proximal lobe was left uncovered.After discussion with the physician, there is no plan for intervention.The patient was on eliquis and aspirin medication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN FLX? Pro
Common Device Name
system, appendage closure, left atrial
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway H91 Y 868
EI   H91 Y868
Manufacturer Contact
mary crawford
4100 hamline ave n
arden hills, MN 55112
+1(651)440-0979
MDR Report Key50734763
Report Number2124215-2026-48725
Device Sequence Number10591492
Product Code NGV
UDI-Device Identifier00191506004590
UDI-Public(01)00191506004590(17)290318(10)38871621
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM635WU60240
Device Catalogue NumberM635WU60240
Device Lot Number0038871621
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2026
Initial Report FDA Received Date09/11/2026
Date Device Manufactured03/19/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age84 YR
Patient SexMale
-
-