Model Number PED2-500-25
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report of pipeline failure to open in the middle section.The patient was undergoing surgery for treatment of a saccular, unruptured, tandem aneurysm in the c6 and c7 segments of the internal carotid artery.It had a max diameter of 5 mm and a 4 mm neck diameter.The landing zone was 3.2 mm distally and 5.1 mm proximally.It was noted the patient's vessel tortuosity was moderate.Dual antiplatelet therapy (dapt) was administered.The angiographic result post procedure showed flow-diverting stent implantation for internal carotid artery aneurysm it was reported that during treatment, a pipeline ped2-500-25 was used.The tip of the stent was successfully opened and positioned in the middle cerebral artery.During continued deployment, when the stent had opened to the midsection, it failed to expand.Multiple attempts to push and manipulate the device were made, but the stent still did not open.Deployment was then continued, and it was also able to open near the terminal segment; however, the midsection still could not open.It was therefore replaced with a ped2-500-20.The pipeline was not used for an indication that is not approved (off-label).The pipeline and accessory devices were prepared as indicated in the instructions for use (ifu).There were no patient symptoms or complications associated with this event.Ancillary devices include a ruikangtong 6f-115 guide catheter, a phenom 27 microcatheter, and a shenruida 200cm guidewire.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
Micro Therapeutics, Inc. dba ev3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
Micro Therapeutics, Inc. dba ev3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
+1(763)526-5677
MDR Report Key50734992
Report Number2029214-2026-01827
Device Sequence Number13416943
Product Code OUT
UDI-Device Identifier00763000449575
UDI-Public(01)00763000449575(17)290119(10)D089406
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPED2-500-25
Device Catalogue NumberPED2-500-25
Device Lot NumberD089406
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/11/2026
Date Device Manufactured01/20/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight121 LB
-
-