• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAPRO?; catheter, coronary, atherectomy

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAPRO?; catheter, coronary, atherectomy Back to Search Results
Model Number H749394671250
Medical Device Problem Codes Failure to Advance (2524); Device Damaged by Another Device (2915); Output Problem (3005)
Health Effect - Clinical Code Perforation (2001)
Date of Event 07/31/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the wire detached and a perforation occurred requiring additional surgery.The 15mm long and 95% stenosed target lesion was located in the severely tortuous and severely calcified mid left anterior descending artery with a vessel diameter of 3.0 mm.A 1.25mm rotapro and rotawire drive were selected for use.During the procedure, difficulty crossing the lesion was experienced.Seventeen ablations were performed at a set speed of 190,000 rpm for 15 seconds each.The physician noted reductions in speed ranging from 2,000 to 10,000 rpm during ablation.A perforation was observed on the opposite side of the calcified nodule and upon withdrawal of the devices it was noted that the rotawire had separated a few centimeters proximal to the radiopaque marker.It was considered possible that the separation was due to burr contact.Bypass grafting of the proximal lad was performed to treat the perforation.No device fragments were left inside the body.Following surgery, the patient was weaned from mechanical ventilation, resumed oral intake, and is currently undergoing rehabilitation.
 
Additional Manufacturer Narrative
A4 weight: 44.7kg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO?
Common Device Name
catheter, coronary, atherectomy
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
business and technology park,
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
+1(612)875-7234
MDR Report Key50735305
Report Number2124215-2026-48729
Device Sequence Number10662047
Product Code MCX
UDI-Device Identifier08714729999560
UDI-Public(01)08714729999560(17)270811(10)37176890
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH749394671250
Device Catalogue NumberH749394671250
Device Lot Number0037176890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/11/2026
Date Device Manufactured08/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
Patient SexMale
-
-