| Model Number |
M635WU60270 |
| Medical Device Problem Codes |
Migration (4003); Failure to Seal (4070)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Reported via patient call center
it was reported that device migration occurred and the implant did not seal.A left atrial appendage closure (laac) procedure was performed, and a 27mm watchman flx pro laa closure device was implanted on (b)(6)-2024.The patient called the watchman patient call center and reported the device had become unrooted and rotated, creating a large crevice in the roof of the appendage.The patient was in for transesophageal echocardiography (tee) imaging prior to a pulmonary vein isolation (pvi) procedure when the suspected leak was discovered.The patient reports imaging at the first follow up was normal.Imaging at the one year follow up showed a small leak.Computed tomography (ct) imaging has been scheduled for 14-sep-2026, and a follow up physician visit on (b)(6)-2026.
|
| |
|
Event or Problem Description
|
|
Reported via patient call center
it was reported that device migration occurred and the implant did not seal.A left atrial appendage closure (laac) procedure was performed, and a 27mm watchman flx pro laa closure device was implanted on (b)(6) 2024.The patient called the watchman patient call center and reported the device had become unrooted and rotated, creating a large crevice in the roof of the appendage.The patient was in for transesophageal echocardiography (tee) imaging prior to a pulmonary vein isolation (pvi) procedure when the suspected leak was discovered.The patient reports imaging at the first follow up was normal.Imaging at the one year follow up showed a small leak.Computed tomography (ct) imaging has been scheduled for (b)(6)2026, and a follow up physician visit on 18-sep-2026.It was further reported, after speaking with the implanting physician, that there was no leak found at the one year follow up appointment.The device however did not still meet all of the position, anchor, size and seal (pass) criteria at that time.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: describe event or problem - updated
h6 device code: failure to seal a050705 - removed per additional information.
|
| |
|