U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba Sprint Quattro Secure MRI SureScan; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba Sprint Quattro Secure MRI SureScan; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 6947M62
Medical Device Problem Code Defibrillation/Stimulation Problem (1573)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that short circuit protection was triggered during an episode of high voltage therapy resulting in less than the programmed high voltage energy being delivered and an alert was triggered.The implantable cardioverter defibrillator (icd was explanted and replaced and the right ventricular (rv) lead was capped and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: product id dtpa2qq; (b)(6); product type: 0236-crt-d; implant date (b)(6) 2023; explant date (b)(6) 2026.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sprint Quattro Secure MRI SureScan
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50736992
Report Number2649622-2026-27068
Device Sequence Number16095787
Product Code NVY
UDI-Device Identifier00643169356627
UDI-Public(01)00643169356627(17)181214
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/14/2018
Device Model Number6947M62
Device Catalogue Number6947M62
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2026
Initial Report FDA Received Date09/11/2026
Date Device Manufactured01/25/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
Patient SexMale
-
-