| Model Number |
AG740M |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Obstruction/Occlusion (2422); Thromboembolism (2654)
|
| Date of Event |
08/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Artegraft implanted on (b)(6) 2026 was closed and an embolectomy procedure to be performed.Email provided from dr.(b)(6) ;
"the thrombectomy appears to have been successful (so far).My findings are as follows:
since i had no prior experience performing a thrombectomy on an artegraft, i removed the clot using a modified thrombosuction technique.I attached a 12 french catheter to the suction source (using an adapter) and aspirated the entire graft.This worked well.However, a small, adherent piece of clot remained at the arterial anastomosis; i removed this using a red fogarty catheter.A follow-up angiogram showed no residual clot (though there was a stenosis at the venous anastomosis, which was treated with a 10 mm pta balloon).I did notice that the artegraft prosthesis appears very robust.The (pseudo)intima does not detach easily, making it possible to clean the graft thoroughly.It is sturdy enough to allow for a standard thrombectomy using a fogarty balloon.It also appears strong enough to withstand radiological intervention for an obstruction.The wall is sufficiently robust to allow for closure of the arteriotomy.".
|
| |
|
Additional Manufacturer Narrative
|
|
The graft was not received for evaluation.A review of the dhr could not be completed as no lot information was received.There was no information or trend identified that indicates that the issue is graft related.A root cause for this complaint could not be conclusively determined, but is most likely patient specific.Based on the documentation and complaint history review, we do not believe there is a systemic issue with these devices.No further actions were determined to be necessary at this time; all product quality and clinical issues will continue to be monitored within quality assurance trending.No related ncmr/capa was identified.Lemaitre current risk controls were determined to be sufficient at this time.
|
| |
|
Search Alerts/Recalls
|