Model Number AG740M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Obstruction/Occlusion (2422); Thromboembolism (2654)
Date of Event 08/13/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Artegraft implanted on (b)(6) 2026 was closed and an embolectomy procedure to be performed.Email provided from dr.(b)(6) ; "the thrombectomy appears to have been successful (so far).My findings are as follows: since i had no prior experience performing a thrombectomy on an artegraft, i removed the clot using a modified thrombosuction technique.I attached a 12 french catheter to the suction source (using an adapter) and aspirated the entire graft.This worked well.However, a small, adherent piece of clot remained at the arterial anastomosis; i removed this using a red fogarty catheter.A follow-up angiogram showed no residual clot (though there was a stenosis at the venous anastomosis, which was treated with a 10 mm pta balloon).I did notice that the artegraft prosthesis appears very robust.The (pseudo)intima does not detach easily, making it possible to clean the graft thoroughly.It is sturdy enough to allow for a standard thrombectomy using a fogarty balloon.It also appears strong enough to withstand radiological intervention for an obstruction.The wall is sufficiently robust to allow for closure of the arteriotomy.".
 
Additional Manufacturer Narrative
The graft was not received for evaluation.A review of the dhr could not be completed as no lot information was received.There was no information or trend identified that indicates that the issue is graft related.A root cause for this complaint could not be conclusively determined, but is most likely patient specific.Based on the documentation and complaint history review, we do not believe there is a systemic issue with these devices.No further actions were determined to be necessary at this time; all product quality and clinical issues will continue to be monitored within quality assurance trending.No related ncmr/capa was identified.Lemaitre current risk controls were determined to be sufficient at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Artegraft
Common Device Name
Collagen Vascular Graft
Manufacturer (Section D)
LeMaitre Vascular, Inc.
206 north center drive
north brunswick NJ 08902
Manufacturer (Section G)
LeMaitre Vascular, Inc.
206 north center drive
north brunswick 08902 -424
Manufacturer Contact
monte nelson
206 north center drive
north brunswick 08902--424
+1(732)422-8333
MDR Report Key50737215
Report Number2247686-2026-00018
Device Sequence Number10719926
Product Code LXA
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
N16837
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberAG740M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-