It was clarified that the following previously reported results are repeat results for the same reported patient:\
184 pmol/l, accompanied by a data flag.\
202 pmol/l, accompanied by a data flag.\
208 pmol/l, accompanied by a data flag.\
241 pmol/l, accompanied by a data flag.\
877 pmol/l, accompanied by a data flag.\
\
the alarm trace data (b)(6)2026) identified 11 sample clot and 25 sample short alarms, which is indicative of poor pre-analytical handling.\
\
the investigation determined the issue was consistent with pre-analytical handling issues at the customer site.Correct pre-analytic sample handling is within the customer's responsibility.
|