• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Roche Diagnostics Elecsys Estradiol III; Estradiol test system

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Roche Diagnostics Elecsys Estradiol III; Estradiol test system Back to Search Results
Catalog Number 09755659190
Medical Device Problem Codes Incorrect Measurement (1383); High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was clarified that the following previously reported results are repeat results for the same reported patient:\ 184 pmol/l, accompanied by a data flag.\ 202 pmol/l, accompanied by a data flag.\ 208 pmol/l, accompanied by a data flag.\ 241 pmol/l, accompanied by a data flag.\ 877 pmol/l, accompanied by a data flag.\ \ the alarm trace data (b)(6)2026) identified 11 sample clot and 25 sample short alarms, which is indicative of poor pre-analytical handling.\ \ the investigation determined the issue was consistent with pre-analytical handling issues at the customer site.Correct pre-analytic sample handling is within the customer's responsibility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Elecsys Estradiol III
Common Device Name
Estradiol test system
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Roche Diagnostics GmbH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50737226
Report Number1823260-2026-03734
Device Sequence Number13586388
Product Code CHP
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number09755659190
Device Lot Number931204
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/20/2026
Supplement Date Received by Manufacturer09/18/2026
Initial Report FDA Received Date09/11/2026
Supplement Report FDA Received Date10/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-