Model Number OTV-S200
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the device was not functioning during maintenance involving an olympus video system center.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was not returned to olympus for inspection however, the technical assistance center (tac) confirmed the reported failure.A definitive root cause could not be identified.Based on the results of the investigation, it is likely that the malfunction was triggered by a bent trigger cable, which caused the device to become inoperative.The customer contacted olympus technical assistance support (tac) via email.During troubleshooting, (b)(6) replaced the damaged trigger cable with the record trigger cable.The customer informed tac of the event.The device will not be returned to olympus for evaluation.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE II VIDEO SYSTEM CENTER
Common Device Name
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50738537
Report Number3002808148-2026-32028
Device Sequence Number10620358
Product Code FET
UDI-Device Identifier04953170386022
UDI-Public04953170386022
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K200542
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOTV-S200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/14/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1