• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Philips Medical Systems Patient Information Center iX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Philips Medical Systems Patient Information Center iX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) Back to Search Results
Model Number Patient Information Center iX
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/15/2026
Type of Reportable Event Death
Event or Problem Description
It was reported that there was no alleged harm or equipment malfunction, and the customer was seeking alarm data on a deceased patient.Biomed was guided in the use of the clinical audit trail and discovered that leads were off of the patient at 07:55:56, which persisted until 08:52.Additionally, alarms and reminders were working as expected.Reminders were seen at roughly three-minute intervals and alarms were silenced concurrently with other alarms.There is insufficient information to determine whether a device malfunction occurred or whether the device was related to the reported event; therefore, good faith efforts are being performed.
 
Additional Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Patient Information Center iX
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medical Systems
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
Philips Medical Systems
3000 minuteman rd
andover MA 01810
Manufacturer Contact
deborah currlin
3000 minuteman rd
andover, MA 01810
+1(617)245-5900
MDR Report Key50741641
Report Number1218950-2026-100625
Device Sequence Number8862354
Product Code MHX
UDI-Device Identifier00884838048645
UDI-Public(01)00884838048645(10)B.02.12
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K153702
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPatient Information Center iX
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-