It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) and electrode exhibited elevated system impedance measurement of 120 ohms.A 10j shock was performed, and a delivered shock impedance of 124 ohms was noted.It was noted that the s-icd was scheduled for a routine battery replacement, and it was not possible to retrieve device data because the s-icd was at end of life (eol).Technical services (ts) reviewed next steps for the electrode included surgical revision or defibrillation threshold (dft) testing.If dfts were unsuccessful, the options were to surgically revise or consider a different implantable system, as it was possible this may be a higher dft patient.Fluoroscopic imaging may also be used to look for adipose tissue.Additional information received reported that the electrode had been implanted in a suboptimal location, to the left of the sternum.Subsequently, the s-icd was explanted, the electrode was surgically abandoned, and a new s-icd system was implanted.No additional adverse patient effects were reported.The s-icd was retained by the hospital and will not be returned.
|