It was reported that electrogram (egm) review was requested for this cardiac resynchronization therapy defibrillator (crt-d) due to observed pauses less than two seconds.Upon review, each pause was preceded by an atrial fibrillation (af) beat.When atrial flutter response (afr) is triggered, post-ventricular atrial refractory period (pvarp) is extended and the ventricular pace (vp) cannot track the p-wave.This appears related to farfield oversensing marked as premature atrial contraction (pac), just prior.Technical services (ts) discussed troubleshooting options and programming considerations, recommending extending the blanking from smart to fixed and turning off afr.Additional information received reported that the recommended programming changes were completed.This crt-d remains in service, and will continue to be monitored.No additional adverse patient effects were reported.
|