Effective (b)(6) 2026, mdr-related activities and responsibilities were transferred to pentax medical corporation as part of a corporate restructuring.This report is being submitted by pentax medical corporation.In accordance with 21 cfr 803.3(m), because this corporate restructuring does not affect the manufacturing site, the fda registration number previously assigned for this site (#9610877) continues to be used.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cmos module with drive pcb signs of fluid droplets on the video image.Based on the result, we concluded that it was caused due to the moisture condensation in the cmos module with drive pcb.In addition, our technician confirmed that the control body fluid damage, the mechanical base plate fluid damage, the insertion flexible tube buckled, the control body corroded, the mechanical base plate corroded, the operation channel (primary) leak, the operation channel (primary) cut, the angle wire hard to move, the led cover glass signs of moisture or fluid under and the cover glass; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|