• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. CS100 Intra-Aortic Balloon Pump, English, 110V; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. CS100 Intra-Aortic Balloon Pump, English, 110V; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0998-00-3013-53
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer that the cs100 intra-aortic balloon pump (iabp) gave an error: maintenance required, code 9.It is unknown when the malfunction occurred.No patient involved and no harm is reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that they were not able to reproduce the error occurred.As per manual, they removed and replaced the fiber optic module as required.Completed repair successfully.Product passed all functional and safety tests per manufacturer specifications.
 
Additional Manufacturer Narrative
Cs100 upgraded to cs300.It gets labeled with the stickers showing upgraded with conversion kit.No product details are changed.Upon completion of the investigation into this event a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 Intra-Aortic Balloon Pump, English, 110V
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
Datascope Corp.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
ewelina chorlowska
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key50749858
Report Number2249723-2026-0004263
Device Sequence Number10221752
Product Code DSP
UDI-Device Identifier10607567107394
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K031636
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number0998-00-3013-53
Device Catalogue Number0998-00-3013-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/02/2015
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-