| Model Number |
0998-00-3013-53 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that the cs100 intra-aortic balloon pump (iabp) gave an error: maintenance required, code 9.It is unknown when the malfunction occurred.No patient involved and no harm is reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that they were not able to reproduce the error occurred.As per manual, they removed and replaced the fiber optic module as required.Completed repair successfully.Product passed all functional and safety tests per manufacturer specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Cs100 upgraded to cs300.It gets labeled with the stickers showing upgraded with conversion kit.No product details are changed.Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|