| Model Number |
D232 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data was sent for engineering analysis.Analysis showed that the battery voltage is dropping rapidly and unpredictably triggering the low voltage anomaly.Device replacement was suggested.To date, this device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data was sent for engineering analysis.Analysis showed that the battery voltage is dropping rapidly and unpredictably triggering the low voltage anomaly.Device replacement was suggested.To date, this device remains in service.No adverse patient effects were reported.Additional information received indicates that the device was explanted and a new device was successfully placed.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report was created to capture additional information in b5 event description and h6 impact codes.
|
| |
|