• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Maquet Cardiopulmonary GmbH Cardiohelp; Console, Heart-Lung Machine, Cardiopulmonary Bypass

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Maquet Cardiopulmonary GmbH Cardiohelp; Console, Heart-Lung Machine, Cardiopulmonary Bypass Back to Search Results
Model Number Cardiohelp
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
The event was occurred in france during installation.It was reported that the cardio help displayed the error message "case 2 fan defective." on (b)(6) 2026, on the provided logfiles of the cardio help device it was found that drive fan 2, housing fan 1 and housing fan 2 defective.No harm to any person has been reported.As both fans of the same type are affected, improper cooling of the product can occurred, therefore a report is required.Complaint #(b)(4).
 
Additional Manufacturer Narrative
A getinge technician was sent for investigation.Technician confirms that the faulty fan is one of the 2 located on the engine on the outside, however both fans were replaced as a precaution.The technician performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.During the review of the logfiles, it was found that drive fan 1 and drive fan 2 were defective, a further clarification is pending from technician.The investigation is ongoing.A follow up medwatch will be submitted when further information becomes available.Note: this event occurred on the france market.It is a significantly similar device to "base unit, cardiohelp¿ which is sold in the usa under premarket submission number k133598.For section d4 all available identifying information is for the out of the us device, subjected to this report.Therefore, no gudid information exists.Furthermore, udi information (primary di number) provided in d4 is for cardiohelp with catalog number 701048012.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Cardiohelp
Common Device Name
Console, Heart-Lung Machine, Cardiopulmonary Bypass
Manufacturer (Section D)
Maquet Cardiopulmonary GmbH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
Julia Kapfenberger
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key50750812
Report Number8010762-2026-0000552
Device Sequence Number8861680
Product Code DTQ
UDI-Device Identifier04037691658384
UDI-Public(01)04037691658384
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K133598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCardiohelp
Device Catalogue Number701048012
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/14/2026
Date Device Manufactured08/12/2016
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-