The event was occurred in france during installation.It was reported that the cardio help displayed the error message "case 2 fan defective." on (b)(6) 2026, on the provided logfiles of the cardio help device it was found that drive fan 2, housing fan 1 and housing fan 2 defective.No harm to any person has been reported.As both fans of the same type are affected, improper cooling of the product can occurred, therefore a report is required.Complaint #(b)(4).
|
|
A getinge technician was sent for investigation.Technician confirms that the faulty fan is one of the 2 located on the engine on the outside, however both fans were replaced as a precaution.The technician performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.During the review of the logfiles, it was found that drive fan 1 and drive fan 2 were defective, a further clarification is pending from technician.The investigation is ongoing.A follow up medwatch will be submitted when further information becomes available.Note: this event occurred on the france market.It is a significantly similar device to "base unit, cardiohelp¿ which is sold in the usa under premarket submission number k133598.For section d4 all available identifying information is for the out of the us device, subjected to this report.Therefore, no gudid information exists.Furthermore, udi information (primary di number) provided in d4 is for cardiohelp with catalog number 701048012.
|