| Model Number |
0684-00-0568-01U |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Device Sensing Problem (2917); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an emergency support program call, the customer reported a fiber-optic sensor failure alarm following an attempt by a provider to reposition the sensation plus 7.5fr.40cc intra-aortic balloon (iab) catheter at the bedside.Augmentation was present; however, the arterial pressure displayed as 224/219 mmhg.The customer was unable to aspirate the inner lumen, and the lumen was considered clotted; guidance was provided to cap and label the inner lumen and discontinue its use.The customer was guided through connecting a radial arterial line to the pump, after which appropriate arterial pressure waveforms and values were obtained.The customer was instructed to retain the iab catheter for evaluation and return it using a biohazard kit.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6)
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|